Advertisement

Alert
Alert
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual

Alerts

All alerts are red. When lit, the alerts indicate the following:
: An alert condition exists.
Alert
: Amplitude is out of range.
Vibration
Connector not plugged into its receptacle on the console rear panel.
Footswitch:
Connector not plugged into its receptacle on the console.
Handpiece:
Cooling water problem in the system.
Cooling Water:

Lap Mode, Prime, Test, and Status Buttons

To activate Lap Mode, Prime, or Status, press the button once; to deactivate the
function, press the button again. To activate Test, press the button once; it
automatically deactivates when the test is complete.
Selects the Laparoscopic mode of operation—the system provides no
Lap Mode:
suction to the handpiece until you activate the Vibration pedal.
When you activate the Lap Mode, the system turns on the Lap On indicator light.
Automatically increases irrigation rate to 25 to 30 cc/min to pump
Prime:
irrigation fluid to the tip. A timer turns Prime off after approximately one minute.
Verifies the handpiece is working properly by automatically increasing tip
Test:
amplitude to 100%, then decreasing it to 0%—all within 4 seconds.
Toggles the status between Standby and Run.
Status:

Displays

Green light emitting diodes (LEDs) show user adjusted setpoints for Aspiration,
Irrigation, and Amplitude:
Not Activated: One LED illuminates to show the setpoint value
Activated: All LEDs up to and including the setpoint value light up.
For TissueSelect, "Std" (standard) is the first setpoint value; "++++" is the highest
setpoint value.
Control Panel
4-5