Table 7-2.
Suction pump voltage steps
CUSA EXcel Ultrasonic Surgical Aspiration System Service Manual
Suction pump control results from the same kind of operation as irrigation pump
control. The 4 bit output signal range for the suction pump is approximately 0 to
6V DC. The following table shows the suction steps:
suction step
desired voltage
default
4.4
max
6
Machine Control Board
+5v_sp = 6.15v
setpoint
theoret v
1
0.40
2
0.80
3
1.20
4
1.61
5
2.00
6
2.41
7
2.81
8
3.23
9
3.63
10
4.03
11
4.43
12
4.84
13
5.23
14
5.64
15
6.03