CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Re-install the Left Cover
Position the bottom of the cover on the left side of the console and tilted
1.
slightly outward.
Push the top of the cover inward toward the console. When the cover is in
2.
place, push the left side latch toward the rear of the console.
Replace the left side lock screw (M5 Phillips), located under the console top
3.
near the left rear corner.
Re-install the Right Cover
Position the bottom of the cover on the right side of the console and tilted
1.
slightly outward. Be sure that the cooling water line service loop fits into the
console side.
Push the top of the cover inward toward the console. When the cover is in
2.
place, push the right side latch toward the rear of the console.
Replace the right side lock screw (M5 Phillips), located under the console top
3.
near the right rear corner.
Replace the cooling water reservoir.
4.
Re-install the Rear Cover
Position the bottom of the cover on the rear of the console and tilted slightly
1.
outward. Allow it to slide downward and to hook into the console.
Hold the top of the cover against the console and replace the two screws (M5
2.
Phillips).
Reconnect the footswitch connector to the receptacle on the rear panel.
3.
Console Covers
11-5