Figure 11-22.
Disconnecting cables from the suction
pump to the suction pump controller
card
Table 11-5.
Suction pump cable connections to the
controller card
CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual
Disconnect the cables from the suction pump to the suction pump controller
3.
card:
Connector
Wire Color
3-pin, white
red, blue, yellow
7-pin, black
yellow, brown, green, red, orange
*
The jumper wire occupies pins 6 and 7 on the connector.
Pull the strain relief bushing for suction pump motor wires away from the
4.
manifold bracket.
Notice
If you are installing the Noise Reduction Kit (NRK) to the system, refer to the
following addendum for instructions:
Addendum to CUSA EXcel
Installing the Noise Reduction Kit (Document # 510I000IE)
Perform the procedures in this addendum in place of the instructions described
on pages 11-32 through 11-35 in this service manual.
Suction Pump and Controller Card
*
®
and CUSA EXcel
®
-8 Service Manuals:
11-31