Speaker; Volume Control; Footswitch Connector Receptacle; Ac Main Switch - Integra Cusa Excel Service Manual

Ultrasonic surgical aspirator
Hide thumbs Also See for Cusa Excel:
Table of Contents

Advertisement

CUSA EXcel Ultrasonic Surgical Aspirator System Service Manual

Speaker

The speaker sounds an audible tone when an alert condition exists and when
vibration is active.

Volume Control

This potentiometer allows you to adjust the active vibration audible tone. You
cannot adjust the alert tone.

Footswitch Connector Receptacle

This receptacle allows you to connect the footswitch to the console.

AC Main Switch

This switch controls the flow of AC power to the CUSA EXcel system. It does not
activate the system; instead, it allows power to flow from the wall receptacle to
the System Power Switch.
Once the AC Main Switch is On ( | ), you can leave it On even when you unplug
the system from the wall receptacle.

Fuses

These protect the system from electrical overloads. Refer to Section 11,
Replacement Procedures for instructions on replacing the fuses.

Footswitch

The footswitch has two pedals in a base, and allows you to activate two functions:
Ultrasonics/vibration: Right pedal (orange), when pressed, activates vibration at
the tip. When you release the pedal, vibration stops.
Fast Flush: Left pedal (blue), when pressed, increases the irrigation rate to
25 to 30 cc/min. It also closes the Suction Pinch Valve, stopping suction. When
you release the pedal, irrigation returns to the setpoint value and suction
resumes.
Console Body Rear Panel
4-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents