D-ONE – Operating instructions V01
5
Maintenance
W
ARNING
D-ONE maintenance should be carried out on a clean and dust free
workplace. Always turn off power and disconnect D-ONE from the mains and
wear gloves when carrying out maintenance work.
D-ONE modules are precision instruments, therefore a proper maintenance routine must
be followed to ensure safe and reliable operation. Cleaning is recommended if D-ONE has
been contaminated or if it has come in contact with corrosive liquids.
5.1

Maintenance schedule

5.1.1

Daily

• Inspect D-ONE for visual damages.
• Clean the outer surface of D-ONE (see 5.2).
5.1.2

Periodical

• In case D-ONE is in daily use, perform a leak test every 3 months (see 5.4).
5.1.3

Yearly Maintenance

• Send it for calibration to INTEGRA at least once a year (see 5.5).
• If you operate the battery beyond the recommended 3 year period, visually check the
battery for signs of damage, e.g. discoloration, unexpected stains, shrinking of the
tube wrapping.
5.2

Cleaning

The materials used on the exterior of D-ONE support regular cleaning intervals. Clean the
external components with a lint-free cloth lightly soaked with mild soap solution in distilled
water or with a 70 % dilution of Isopropyl or Ethanol. Never use Acetone or other solvents.
W
ARNING
Do not immerse the entire D-ONE into a cleaning solution or spray cleaning
solution directly onto the exterior body of D-ONE as this can potentially
damage internal electronics.
If liquid ever enters the internals of D-ONE, please contact your service
technician.
C
AUTION
D-ONE can only be completely disassembled by trained service personnel.
www.integra-biosciences.com