Verification Tests; Zero Offset Verification - Agilent Technologies 34410A Service Manual

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

3
Calibration Procedures

Verification Tests

Zero Offset Verification

This procedure is used to check the zero offset performance of the
instrument. Verification checks are only performed for those functions and
ranges with unique offset calibration constants. Measurements are checked
for each function and range as described in the procedure on the next
page.
Zero Offset Verification Procedure (34410A/11A)
1 Make sure you have read
2 Install the 34172B Calibration Short (or equivalent) on the front panel
3 Select each function and range in the order shown in the table below.
4 Install the 34172B Calibration Short (or equivalent) on the rear panel
5 Select each function and range in the order shown in the table below.
Zero Offset Verification Procedure (L4411A only)
1 Make sure you have read
2 Install the 34172B Calibration Short (or equivalent) on the input
3 Select each function and range in the order shown in the table below.
64
input terminals. Select the Front input terminals with the Front/Rear
switch.
Make a measurement and return the result. Compare measurement
results to the appropriate test limits shown in the table on the next
page.
input terminals. Select the Rear input terminals with the Front/Rear
switch.
Make a measurement and return the result. Compare measurement
results to the appropriate test limits shown in the table below.
terminals.
Make a measurement and return the result. Compare measurement
results to the appropriate test limits shown in the table on the next
page.
"Test Considerations"
"Test Considerations"
34410A/11A/L4411A Service Guide
on page 63.
on page 63.