When the pass/fail test function is enabled (refer to the :MASK:ENABle command), this command is
invalid.
The waveform auto setting function requires that the frequency of sine waveform is no less than 20 Hz
the duty cycle is greater than 1%, and the amplitude is at least 20 mVpp. Otherwise, the waveform
auto setting function may be invalid.

:CLEar

Syntax
:CLEar
Description
Clears all the waveforms on the screen. If the oscilloscope is in the
cleared, new waveform data will be displayed. This command functions the same as the CLEAR key on the
front panel.

:RUN

:STOP

Syntax
:RUN
:STOP
Description
The :RUN command starts the oscilloscope and the :STOP command stops the oscilloscope. This command
functions the same as the Run/Stop key on the front panel.
Remarks
If the oscilloscope is in a single trigger mode (refer to the :SINGle command), you can use the :RUN
command to set the trigger mode of the oscilloscope to "Auto", and keep the oscilloscope in the "Run" state.
If the oscilloscope is in a single trigger mode (refer to the :SINGle command) and in WAIT state, you can
use the :STOP command to set the trigger mode of the oscilloscope to "Auto", and keep the oscilloscope in
the "Stop" state.

:SINGle

Syntax
:SINGle
Description
Sets the trigger mode of the oscilloscope to "Single". This command functions the same as the Single key
on the front panel.
Remarks
In the single trigger mode, the oscilloscope performs a single trigger when the trigger conditions are
met and then it stops. You can use the :TFORce command to generate a trigger by force.
If the oscilloscope is in a single trigger mode, you can use the :RUN command to set the trigger mode
of the oscilloscope to "Auto", and keep the oscilloscope in the "Run" state.
If the oscilloscope is in a single trigger mode and in WAIT state, you can use the :STOP command to
set the trigger mode of the oscilloscope to "Auto", and keep the oscilloscope in the "Stop" state.
2-2
Chapter 2 Command System
RUN
state, after the waveform data is
DS4000E Programming Guide
,