RIGOL

:WAVeform:END

Syntax
:WAVeform:END
Description
Stops the waveform data reading.
Remarks
This command is only applicable to reading waveform data in the internal memory (refer to
the :WAVeform:MODE command).

:WAVeform:FORMat

Syntax
:WAVeform:FORMat <format>
:WAVeform:FORMat?
Description
Sets or queries the return format of the waveform data.
Parameter
Name
Type
<format>
Discrete
Remarks
WORD: Each waveform point corresponds to 2-byte (16-bit) waveform data. The lower 8 bits are valid
and the higher 8 bits are 0.
BYTE: Each waveform point corresponds to 1-byte (8-bit) waveform data.
ASCii: The query directly returns the actual voltage value of each waveform point in scientific notation;
and the voltage values are separated by commas.
Return Format
The query returns WORD, BYTE, or ASC.
2-242
Range
{WORD|BYTE|ASCii}
Chapter 2 Command System
TMC
Data
Block
Header
Default
BYTE
DS4000E Programming Guide