Measure:pduty - Rigol DS4000E Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for DS4000E Series:
Table of Contents

Advertisement

RIGOL
When [<chan>] is omitted, the command, by default, applies to the currently set measurement source
(refer to the :MEASure:SOURce command).
Return Format
The query returns the average value, current value, standard deviation, maximum value, or minimum value
in scientific notation. When the measurement results are invalid, the query returns 9.9e37.

:MEASure:PDUTy

Syntax 1
:MEASure:PDUTy [<chan>]
Description 1
Measures the positive duty cycle of the specified source channel.
Syntax 2
:MEASure:PDUTy? [<chan>]
Description 2
Measures and queries the positive duty cycle of the specified source channel.
Parameter
Name
Type
<chan>
Discrete
Remarks
Before making measurements, turn on the channels to be measured.
When the math operation type is FFT (refer to the :CALCulate:MODE command), <chan> cannot be
set to "MATH".
When [<chan>] is omitted, the command, by default, applies to the currently set measurement source
(refer to the :MEASure:SOURce command).
Return Format
The query returns the measurement results in scientific notation. When the measurement results are
invalid, the query returns 9.9e37.
:MEASure:PDUTy:SAVerage
:MEASure:PDUTy:SCURrent
:MEASure:PDUTy:SDEViation
:MEASure:PDUTy:SMAXimum
:MEASure:PDUTy:SMINimum
Syntax 1
:MEASure:PDUTy:SAVerage [<chan>]
:MEASure:PDUTy:SCURrent [<chan>]
:MEASure:PDUTy:SDEViation [<chan>]
:MEASure:PDUTy:SMAXimum [<chan>]
:MEASure:PDUTy:SMINimum [<chan>]
Description 1
Measures the positive duty cycle of the specified source channel and enables the statistic function of the
measurement results.
2-136
Range
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH}
Chapter 2 Command System
Default
--
DS4000E Programming Guide