Chapter 2 Command System

:MEASure:VMIN

Syntax 1
:MEASure:VMIN [<chan>]
Description 1
Measures the minimum amplitude value of the specified source channel.
Syntax 2
:MEASure:VMIN? [<chan>]
Description 2
Measures and queries the minimum amplitude value of the specified source channel. The unit is related to
the current amplitude unit of the specified source channel.
Parameter
Name
Type
<chan>
Discrete
Remarks
Before making measurements, turn on the channels to be measured.
When the math operation type is FFT (refer to the :CALCulate:MODE command), <chan> cannot be
set to "MATH".
When [<chan>] is omitted, the command, by default, applies to the currently set measurement source
(refer to the :MEASure:SOURce command).
Return Format
The query returns the measurement results in scientific notation. When the measurement results are
invalid, the query returns 9.9e37.
:MEASure:VMIN:SAVerage
:MEASure:VMIN:SCURrent
:MEASure:VMIN:SDEViation
:MEASure:VMIN:SMAXimum
:MEASure:VMIN:SMINimum
Syntax 1
:MEASure:VMIN:SAVerage [<chan>]
:MEASure:VMIN:SCURrent [<chan>]
:MEASure:VMIN:SDEViation [<chan>]
:MEASure:VMIN:SMAXimum [<chan>]
:MEASure:VMIN:SMINimum [<chan>]
Description 1
Measures the minimum amplitude value of the specified source channel and enables the statistic function of
the measurement results.
Syntax 2
:MEASure:VMIN:SAVerage? [<chan>]
:MEASure:VMIN:SCURrent? [<chan>]
:MEASure:VMIN:SDEViation? [<chan>]
:MEASure:VMIN:SMAXimum? [<chan>]
:MEASure:VMIN:SMINimum? [<chan>]
DS4000E Programming Guide
Range
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH}
RIGOL
Default
--
2-161