:CALibrate Commands

Command List:

:CALibrate:DATE?

:CALibrate:QUIT

:CALibrate:STARt

:CALibrate:TIME?
:CALibrate:DATE?
Syntax
:CALibrate:DATE?
Description
Queries the date when the last self-calibration was performed.
Return Format
The query returns <year>,<month>,<day>. Wherein, <year>, <month>, and <day> represent year,
month, and day respectively.
:CALibrate:QUIT
Syntax
:CALibrate:QUIT
Description
Stops performing the self-calibration operation.
Remarks
You can enable the oscilloscope to stop performing the self-calibration operation at any time.
:CALibrate:STARt
Syntax
:CALibrate:STARt
Description
Enables the oscilloscope to start executing the self-calibration operation.
Remarks
The self-calibration operation enables the oscilloscope to work in an optimal state and obtain accurate
measurement results.
Before executing the self-calibration operation, properly connect the calibration signal to the external
trigger input channel and four input channels.
During the self-calibration process, most of the keys are disabled.
2-58
Chapter 2 Command System
DS4000E Programming Guide