Trigger:flexray - Rigol DS4000E Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for DS4000E Series:
Table of Contents

Advertisement

Chapter 2 Command System

:TRIGger:FLEXray

Command List:
:TRIGger:FLEXray:BAUD
:TRIGger:FLEXray:LEVel
:TRIGger:FLEXray:SOURce
:TRIGger:FLEXray:WHEN
:TRIGger:FLEXray:BAUD
Syntax
:TRIGger:FLEXray:BAUD <baud>
:TRIGger:FLEXray:BAUD?
Description
Sets or queries the signal rate of FlexRay trigger. The default unit is b/s.
Parameter
Name
Type
<baud>
Discrete
Return Format
The query returns 2500000, 5000000, or 10000000.
:TRIGger:FLEXray:LEVel
Syntax
:TRIGger:FLEXray:LEVel <level>
:TRIGger:FLEXray:LEVel?
Description
Sets or queries the trigger level of FlexRay trigger. The unit is the same as the current amplitude unit of the
source channel.
Parameter
Name
Type
<level>
Real
Return Format
The query returns the trigger level in scientific notation.
DS4000E Programming Guide
Range
{2500000|5000000|10000000}
Range
The range of <level> is from (-6 x VerticalScale - Offset)
to (6 x VerticalScale - Offset) (relative to the vertical
center of the screen).
Wherein, VerticalScale is the vertical scale of the source
channel, refer to the :CHANnel<n>:SCALe command;
Offset is the vertical position of the source channel, refer
to the :CHANnel<n>:OFFSet command.
RIGOL
Default
2500000
Default
0
2-237