Trigger:edge - Rigol DS4000E Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for DS4000E Series:
Table of Contents

Advertisement

Chapter 2 Command System

:TRIGger:EDGe

Command List:
:TRIGger:EDGe:LEVel
:TRIGger:EDGe:SLOPe
:TRIGger:EDGe:SOURce
:TRIGger:EDGe:LEVel
Syntax
:TRIGger:EDGe:LEVel <level>
:TRIGger:EDGe:LEVel?
Description
Sets or queries the trigger level of Edge trigger. The unit is the same as the current amplitude unit of the
source channel.
Parameter
Name
Type
<level>
Real
Remarks
The range of <level> is related to the type of the source channel of Edge trigger.
Analog channel (CH1-CH4
The range of <level> is from (-6 x VerticalScale - Offset) to (6 x VerticalScale - Offset) (relative to the
vertical center of the screen).
Wherein, VerticalScale is the vertical scale of the source channel, refer to the :CHANnel<n>:SCALe
command; Offset is the vertical position of the source channel, refer to the :CHANnel<n>:OFFSet
command.
External trigger channel (EXT or EXT/5)
The range of <level> is from -800 mV to +800 mV (EXT) or from -4 V to +4 V (EXT/5). The unit is the
same as the current amplitude unit of CH1.
AC line
You do not need to set the trigger level.
Return Format
The query returns the trigger level in scientific notation.
:TRIGger:EDGe:SLOPe
Syntax
:TRIGger:EDGe:SLOPe <slope>
:TRIGger:EDGe:SLOPe?
Description
Sets or queries the edge type of Edge trigger.
Parameter
Name
Type
<slope>
Discrete
Remarks
POSitive: indicates that the oscilloscope triggers at the crossing point where the rising edge of the
input signal meets the preset trigger level.
DS4000E Programming Guide
Range
Refer to Remarks
)
Range
{POSitive|NEGative|RFALl}
RIGOL
Default
0
Default
POSitive