System Commands; System:aoutput - Rigol DS4000E Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for DS4000E Series:
Table of Contents

Advertisement

RIGOL

:SYSTem Commands

Command List:

:SYSTem:AOUTput

:SYSTem:AUToscale
:SYSTem:BEEPer
:SYSTem:DATE
:SYSTem:ERRor[:NEXT]?
:SYSTem:ETIMpedence
:SYSTem:EXPand
:SYSTem:GAMount?
:SYSTem:GPIB
:SYSTem:LANGuage
:SYSTem:OPTion:INSTall
:SYSTem:OPTion:UNINSTall
:SYSTem:PON
:SYSTem:PSTatus
:SYSTem:RAMount?
:SYSTem:RCLOck
:SYSTem:RESet
:SYSTem:SETup
:SYSTem:SSAVer:TIME
:SYSTem:TIME
:SYSTem:UDEVice
:SYSTem:VERSion?
:SYSTem:AOUTput
Syntax
:SYSTem:AOUTput <aoutput>
:SYSTem:AOUTput?
Description
Sets or queries the type of the signal output from the [Trig Out/Calibration] connector on the rear
panel.
Parameter
Name
Type
<aoutput>
Discrete
Remarks
TOUT: indicates that each time the oscilloscope triggers, a signal that can reflect the current capture
rate of the oscilloscope will be output from the [Trig Out/Calibration] connector on the rear panel.
QEDGe: indicates that a fast edge signal whose rise time is 500 ps will be output from the [Trig
Out/Calibration] connector on the rear panel. The signal can be used for the self-calibration of the
oscilloscope.
GND: indicates that a ground level will be output from the [Trig Out/Calibration] connector on the rear
panel.
PFAil: indicates that when a failed waveform is detected by the oscilloscope, a pulse will be output
from the [Trig Out/Calibration] connector on the rear panel.
Return Format
The query returns TOUT, QEDG, GND, or PFA.
2-180
Range
{TOUT|QEDGe|GND|PFAil}
Chapter 2 Command System
Default
TOUT
DS4000E Programming Guide