RIGOL
:TRIGger Commands
Command List:
:TRIGger:COUPling
:TRIGger:HOLDoff
:TRIGger:MODE
:TRIGger:STATus?
:TRIGger:SWEep
:TRIGger:NREJect
:TRIGger:CAN
:TRIGger:EDGe
:TRIGger:IIC
:TRIGger:PATTern
:TRIGger:PULSe
:TRIGger:RUNT
:TRIGger:NEDGe
:TRIGger:RS232
:TRIGger:SLOPe
:TRIGger:SPI
:TRIGger:USB
:TRIGger:VIDeo
:TRIGger:FLEXray
:TRIGger:COUPling
Syntax
:TRIGger:COUPling <couple>
:TRIGger:COUPling?
Description
Sets or queries the trigger coupling type.
Parameter
Name
Type
<couple>
Discrete
Remarks
DC: allows DC and AC components to pass the trigger path.
AC: blocks all the DC components and attenuates signals below 8 Hz.
LFReject: blocks the DC components and rejects the low frequency components below 5 kHz.
HFReject: rejects the high frequency components above 50 kHz.
Trigger coupling is only valid in Edge trigger (refer to the :TRIGger:MODE command).
Return Format
The query returns DC, AC, LFR, or HFR.
2-194
Range
{DC|AC|LFReject|HFReject}
Chapter 2 Command System
Default
DC
DS4000E Programming Guide