Configure process measurement applications
March 2019
Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
2. Scroll to Step 2, set Active Matrix to the matrix you want to use and click Change Matrix.
5.2.4 Set up concentration measurement using a web browser
This section guides you through the tasks required to set up, configure, and implement concentration
measurement.
Enable the concentration measurement application using a web browser
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
The API Referral application must not be enabled on your transmitter.
Procedure
1. Choose Configuration > Transmitter Options and ensure that Volume Flow Type is set to Liquid
Volume.
2. Under Permanent Features, set Concentration Measurement to Enabled and click Apply.
Set reference temperature values for specific gravity using a web browser
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature
for water, then verify the density of water at the configured reference temperature. These values affect
specific gravity measurement.
This requirement applies to the following derived variables:
Specific Gravity
Concentration (Specific Gravity)
Mass Concentration (Specific Gravity)
Volume Concentration (Specific Gravity)
Procedure
1. Choose Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement page is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Under Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change Matrix.
3. Under Step 3, perform the following actions:
70
Configuration and Use Manual
Micro Motion Model 5700 with Ethernet Transmitters
MMI-20029769