Reboot The Transmitter - Emerson Micro Motion 5700 Configuration And Use Manual

With ethernet transmitters
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20029769
7. Verify the transmitter configuration and all safety parameters.
8. Enable write-protection.
15.2.2 Using the USB service port and ProLink III
You can upgrade the transmitter firmware to stay current with development and to take advantage of any
new features.
This procedure is not available over HART. You must use a service port.
CAUTION
If the transmitter is in a hazardous area, do not open the wiring compartment. Contact customer support.
Prerequisites
You must have the firmware upgrade files provided by Emerson.
Procedure
1. Choose Device Tools > Transmitter Software Update.
2. Navigate to the folder containing the firmware upgrade files.
3. Click Update.
Note
If required, the transmitter upgrade procedure automatically includes an upgrade to the core
processor software.
If you chose to reboot the transmitter at a later date, you can reboot it from the display, or you can
power-cycle it.
4. Verify the transmitter configuration and all safety parameters.
5. Enable write-protection.

15.3 Reboot the transmitter

Menu > Service Tools > Reboot Transmitter
Display
ProLink III
Not available
Web browser
Not available
For certain configuration changes to take effect, the transmitter must be rebooted. You must also reboot the
transmitter in order to clear certain status alerts.
Rebooting the transmitter has the same effect as power-cycling the transmitter.
Prerequisites
Follow appropriate procedures to select the appropriate time for rebooting the transmitter. The reboot
typically takes about 10 seconds.