Troubleshooting
March 2019
Cause
A calibration is running.
Recommended actions
Allow the test to complete.
17.12.26 [105] Two-Phase Flow
Alert
Process Aberration
Cause
The line density is outside the user-defined two-phase flow limits.
Recommended actions
1. Check for two-phase flow.
Refer to
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA
2. Verify that the transmitter is configured correctly for the connected sensor.
17.12.27 [110] Frequency Output X Saturated
Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the scaling of the Frequency Output.
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
17.12.28 [111] Frequency Output 1 Fixed
Alert
Output Fixed
Cause
The output is configured to send a constant value. A loop test may be in progress.
Recommended actions
Check to see if a loop test is in process (the output is fixed).
220
Configuration and Use Manual
Micro Motion Model 5700 with Ethernet Transmitters
MMI-20029769
Output.