Emerson Micro Motion 5700 Configuration And Use Manual page 47

With ethernet transmitters
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20029769
ProLink III
Device Tools > Configuration > Process Measurement > Temperature > Temperature Unit
Web browser
Configuration > Process Measurement > Temperature > Temperature Unit
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
Default: °C (Celsius)
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different
communications tools may use different labels for the units.
Unit description
Degrees Celsius
Degrees Fahrenheit
Degrees Rankine
Kelvin
4.6.2 Configure Temperature Damping
Display
Menu > Configuration > Process Measurement > Temperature > Damping
ProLink III
Device Tools > Configuration > Process Measurement > Temperature > Temperature Damping
Web browser
Configuration > Process Measurement > Temperature > Temperature Damping
Temperature Damping controls the amount of damping that will be applied to temperature data from the
sensor. Temperature Damping is not applied to external temperature data.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value
specifies the time period, in seconds, over which the transmitter will spread changes in the process variable.
At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the
change in the actual measured value.
Procedure
Set Temperature Damping to the desired value.
Default: 4.8 seconds
Range: 0.0 to 80 seconds
Note
If a number greater than 80 is entered, it is automatically changed to 80.
Configuration and Use Manual
Display
°C
°F
°R
°K
Configure process measurement
Label
ProLink III
Web browser
°C
°C
°F
°F
°R
°R
°K
°K
March 2019
47

Advertisement