Recommended actions
1. Verify the wiring between the Discrete Output and the valve. The ON signal must open the valve.
Perform a loop test.
2. Change the setting of Discrete Output Polarity.
17.4.3 Valve does not operate
Cause
Discrete Output wiring is incorrect
Discrete Output channel is set up for internal power
Recommended actions
1. Verify the wiring between the Discrete Output and the valve. The ON signal must open the valve.
Perform a loop test.
2. Ensure that the channel is wired and configured for external power.
3. Ensure that the valve and channel are powered.
17.4.4 Batch totals are inaccurate
Cause
Batch target is incorrect
AOC is inaccurate
Flow measurement needs adjustment
Recommended actions
1. Verify that the correct batch preset is active.
2. Verify that the batch target is set correctly.
3. If you have recently changed the batch target, run a few batches and wait for the AOC value to adjust.
4. Repeat AOC calibration.
5. Adjust the fixed AOC value.
6. Compare the batch total to a scale reading.
17.4.5 ts BatchRepeatNotSatis
Cause
AOC is inaccurate
Inconsistent or leaking valve
Process is unstable
Recommended actions
1. Inspect the valve and replace if necessary.
2. Repeat AOC calibration.