Measurement support
March 2019
Postrequisites
View the results and take any appropriate actions.
14.1.2 View SMV test results
After each SMV test, the pass/fail result is displayed automatically. Detailed results are also available for
licensed versions.
Tip
For licensed SMV versions only, when you use either the display or the web browser to view test results, the 20
most recent results are available. If you use ProLink III to view results, you can view results for all tests that are
in the PC database.
View SMV test results using the display
Procedure
For licensed and unlicensed versions of SMV, results of the current test are displayed automatically.
For licensed versions of SMV only, to view results of previous tests for this meter:
a) Choose Menu > Service Tools > Verification & Calibration > Smart Meter Verification > Read SMV
History.
Pass/Fail results of all tests in the transmitter's SMV database are displayed.
b) To view detailed data for an individual test, select it from the list.
View SMV test results using ProLink III
Procedure
For licensed and unlicensed versions of SMV, results of the current test are displayed automatically.
For licensed versions of SMV only, to view results of previous tests for this meter:
a) Choose Device Tools > Diagnostics > Meter Verification > Run Test.
b) In the SMV Test Definition window, click View Previous Test Results.
c) Click Next.
ProLink III displays a report containing details of the most recent test. The report is automatically
saved to the SMV database. You can print or export the report.
d) To view details of previous tests, click View previous test report.
View SMV test results using a web browser
Procedure
For licensed and unlicensed versions of SMV, results of the current test are displayed automatically.
For licensed versions of SMV only, to view results of previous tests for this meter:
a) Choose Service Tools > Smart Meter Verification > Meter Verification History.
b) Under Meter Verification History select + in the row next to the test date you want to view.
160
Configuration and Use Manual
Micro Motion Model 5700 with Ethernet Transmitters
MMI-20029769