Emerson Micro Motion 5700 Configuration And Use Manual page 66

With ethernet transmitters
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configure process measurement applications
March 2019
.xml (ProLink III)
.matrix (4200)
If you are loading an .xml file, you must know the following information for your matrix:
The derived variable that the matrix is designed to calculate
The density unit that the matrix was built with
The temperature unit that the matrix was built with
If you are loading a .matrix file, you must know the derived variable that the matrix is designed to calculate.
Important
All concentration matrices on your transmitter must use the same derived variable.
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the transmitter's SD card. Set Derived Variable before loading
concentration matrices.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.
Restriction
You cannot use ProLink III to load a matrix to the transmitter's SD card. ProLink III loads matrices directly to
one of the transmitter's six slots.
Procedure
1. If you are loading an .xml file, choose Device Tools > Configuration > Process Measurement > Line
Density and set Density Unit to the density unit used by your matrix.
Important
When you load a matrix in one of these formats, if the density unit is not correct, concentration data
will be incorrect. The density units must match at the time of loading. You can change the density unit
after the matrix is loaded.
2. If you are loading an .xml file, choose Device Tools > Configuration > Process Measurement > Line
Temperatureand set Temperature Unit to the temperature unit used by your matrix.
Important
When you load a matrix in one of these formats, if the temperature unit is not correct, concentration
data will be incorrect. The temperature units must match at the time of loading. You can change the
temperature unit after the matrix is loaded.
3. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
4. In Step 1, ensure that the setting of Derived Variable matches the derived variable used by your
matrix. If it does not, change it as required and click Apply.
66
Configuration and Use Manual
Micro Motion Model 5700 with Ethernet Transmitters
MMI-20029769

Advertisement

Table of Contents
loading

Table of Contents