Emerson Micro Motion 5700 Configuration And Use Manual page 219

With ethernet transmitters
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20029769
2. Check the SMV abort code and take appropriate steps.
3. Contact customer service.
17.12.23 [102] Drive Overrange
Alert
Drive Over-Range
Cause
The drive power (current/ voltage) is at its maximum.
Recommended actions
1. Check your process conditions against the values reported by the device.
2. Check for air in the flow tubes, tubes not filled, foreign material in the tubes, coating in the tubes, or
other process problems.
3. Verify that the tubes are full of process fluid.
4. Check the drive gain and the pickoff voltage.
Refer to
Check the drive gain
5. Check the sensor coils for electrical shorts. If you find problems, replace the sensor. .
Refer to
Check for internal electrical
6. Ensure that the sensor orientation is appropriate for your application.
Settling from a two-phase or three-phase fluid can cause this alert even if the flow tubes are full.
17.12.24 [103] Configuration Warning
Alert
Configuration Warning
Cause
The core processor was unable to store the totalizer and inventory values on the last power-down, and must
rely on the saved totals. (Standard core processor only)
Recommended actions
Verify that the transmitter is receiving sufficient power.
If it is not, correct the problem and reboot or power-cycle the transmitter.
If it is, replace the transmitter because the transmitter has an internal power issue.
17.12.25 [104] Calibration in Progress
Alert
Function Check in Progress
Configuration and Use Manual
and
Check the pickoff voltage
problems.
respectively.
Troubleshooting
March 2019
219

Advertisement

Table of Contents
loading