6.7.2 Interaction between Process Variable Fault Action and other
fault actions
The setting of Process Variable Fault Action affects the operation of the mA Outputs, Frequency Outputs,
and Discrete Outputs if the corresponding output fault actions are set to None.
Interaction between Process Variable Fault Action and mA Output Fault Action
If mA Output Fault Action is set to None, the mA Output signal depends on the setting of Process Variable
Fault Action.
If the device detects a fault condition:
1. Process Variable Fault Action is evaluated and applied.
2. mA Output Fault Action is evaluated.
If it is set to None, the output reports the value associated with the setting of Process Variable
Fault Action.
If it is set to any other option, the output performs the specified fault action.
Configuration and Use Manual
Configure advanced options for process measurement
Label
Web browser
Zero
Not a Number
Flow to Zero
None (default)
Description
Flow rate variables go to the value that
represents a flow rate of 0 (zero).
Density is reported as
0
Temperature is reported as
equivalent if other units are used (e.g.,
.
32 °F
Drive gain is reported as measured.
Totalizers stop incrementing.
Process variables are reported as
.
NAN
Drive gain is reported as measured.
Modbus scaled integers are reported as
.
Max Int
Totalizers stop incrementing.
Flow rates are reported as
Other process variables are reported as
measured.
Totalizers stop incrementing.
All process variables are reported as
measured.
Totalizers increment if they are running.