The concentration measurement application and the advanced phase measurement application
cannot be enabled simultaneously.
6. Enable concentration measurement.
a) Choose Menu > Configuration > Process Measurement > Concentration Measurement.
b) Set Enabled/Disabled to Enabled.
Load a concentration matrix from a USB drive using the display
At least one concentration matrix must be loaded into one of the six slots on your transmitter. You can load
up to six matrices into slots. You can also copy matrices to the transmitter's SD card, and load them into slots
at a later time.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.
WARNING
If the transmitter is in a hazardous area, do not use this method to load matrices. This method requires
opening the transmitter's wiring compartment while the transmitter is powered up, and this can cause an
explosion. If the transmitter is in a hazardous area, you must use another method to load matrices.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The
transmitter's SD card and the ProLink III installation include a set of standard concentration matrices. Other
matrices are available from Micro Motion.
Each concentration matrix file must be in .matrix format.
Tip
If you have a custom matrix on another device, you can save it to a file, then load it to the current device.
If you have a matrix file in a different format, you can load it using ProLink III.
The .matrix files must be copied to the root directory of a USB drive.
You must know the derived variable that the matrix is designed to calculate.
Important
All concentration matrices on your transmitter must use the same derived variable.
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the SD card. Set Derived Variable before loading concentration
matrices.
Procedure
1. Choose Menu > Configuration > Process Measurement > Concentration Measurement > Configure
Application and ensure that the setting of Derived Variable matches the derived variable used by your
matrix. If it does not, change it as required and click Apply.
Configuration and Use Manual
Configure process measurement applications
March 2019
61