View And Acknowledge Status Alerts - Emerson Micro Motion 5700 Configuration And Use Manual

With ethernet transmitters
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Transmitter operation
March 2019
12.1.4 Effect of Sensor Flow Direction Arrow on digital
communications
Flow rates on the transmitter display or reported via digital communications are shown as positive or
negative. The sign depends on the interaction between Sensor Flow Direction Arrow and the actual flow
direction.
This interaction affects flow rates shown on the transmitter display, ProLink III, and all other user interfaces.
Actual flow direction
Forward (same direction
as Flow arrow on sensor)
Reverse (opposite from
Flow arrow on sensor)

12.2 View and acknowledge status alerts

The transmitter posts a status alert whenever one of the specified conditions occurs. You can view active
alerts and you can acknowledge alerts. You do not have to acknowledge alerts: The transmitter will perform
normal measurement and reporting functions with unacknowledged alerts.
12.2.1 View and acknowledge alerts using the display
You can view information about all active or unacknowledged alerts, and you can acknowledge alerts.
The display uses the alert banner and the alert symbol ⓘ to provide information about alerts.
Table 12-1: Alert information on display
Display status
Alert banner
Alert symbol ⓘ
If alert security is enabled, the alert banner is never displayed. To view detailed information, you must use the
alert menu: Menu > (i) Alert List.
Note
Certain alerts do not clear until the transmitter is rebooted.
Procedure
If the alert banner appears:
a) Activate Info to view information about the alert.
b) Take appropriate steps to clear the alert.
c) Activate Ack to acknowledge the alert.
146
Setting of Sensor Flow
Direction Arrow
With Arrow
Against Arrow
With Arrow
Against Arrow
Cause
One or more alerts are
active.
One or more alerts are
unacknowledged.
Flow rate value
Transmitter display
Positive (no sign)
Negative
Negative
Positive (no sign)
User action
Resolve the conditions to clear the alert. When the alert is
cleared or acknowledged, the banner will be removed.
Acknowledge the alert. When all alerts are acknowledged,
the alert icon will be removed.
Micro Motion Model 5700 with Ethernet Transmitters
Configuration and Use Manual
MMI-20029769
Digital communications
Positive
Negative
Negative
Positive