Configuration and Use Manual
MMI-20029769
Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using the display
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Menu > Configuration > Process Measurement > Concentration Measurement > Configure
Application.
2. Set Active Matrix to the matrix you want to use.
5.2.3 Set up concentration measurement using ProLink III
This section guides you through the tasks required to set up, configure, and implement concentration
measurement.
Enable the concentration measurement application using ProLink III
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
The concentration measurement application cannot be enabled at the same time as the API referral
application or the Advanced Phase Measurement application. They must be disabled first.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > Flow and ensure that Volume Flow
Type is set to Liquid Volume.
2. Choose Device Tools > Configuration > Transmitter Options.
3. Set Concentration Measurement to Enabled and click Apply.
Load a concentration matrix using ProLink III
At least one concentration matrix must be loaded onto your transmitter. You can load up to six.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The ProLink III
installation includes a set of standard concentration matrices. Other matrices are available from
Micro Motion. The file can be on your computer or in the transmitter's internal memory.
The file must be in one of the formats that ProLink III supports. This includes:
Configuration and Use Manual
Configure process measurement applications
March 2019
65