Configure process measurement applications
March 2019
Option
Description
Disabled (Off) The preset is not available for use. You cannot select it to run a batch. You can always
enable it at a later time.
Preset 1 is always enabled and cannot be disabled.
7. Optional: For two-stage batching, set % Target or Quantity for the following parameters:
Two-stage batching is a batch that is run on a gas and liquid mixture flowing through a pipeline.
% Target
Quantity
Parameter
End Warning
Open Primary
Open Secondary
Close Primary
5.3.3 Configure a DO for valve control
The transmitter runs a batch by opening and closing a valve. You can optionally configure a Discrete Output
(DO) to send the open and close commands to the valve.
Configure a DO for valve control using the display
Prerequisites
Channel B, Channel C, or Channel D must be available for valve control.
The selected channel must be wired to the valve.
The selected channel and the valve must be powered by a power supply, not by the transmitter.
Procedure
1. Verify the wiring between the selected channel and the valve.
2. Configure the selected channel to operate as a Discrete Output.
a) Choose Menu > Configuration > Inputs/Outputs.
b) Set Channel B, Channel C, or Channel D to operate as a Discrete Output.
78
The values for Open Primary, Open Secondary, Close Primary, and End
Warning are each configured as a percent of target.
The values for Open Primary and Open Secondary are each configured as
a quantity at which the valve should open. The values for Close Primary
and End Warning are each configured as a quantity that is subtracted
from the target.
Description
The % target or quantity level that must be reached to alert the operator
that a batch is ending.
The % target or quantity level that must be reached to open the primary
valve.
The % target or quantity level that must be reached to open the secondary
valve.
The % target or quantity level that must be reached to close the primary
valve. When configuring by quantity, the amount is referenced to the end
of the batch and not to the beginning of the batch.