Configure process measurement applications
March 2019
Procedure
1. Choose Device Tools > Configuration > Process Measurement > API Referral.
2. Choose the method to be used to supply temperature data, and perform the required setup.
Option
Description
Digital
A host writes temperature data
communications
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs
group on the ProLink III main window.
The current pressure value is displayed in the External Pressure field. Verify that the value is correct.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For digital communications:
Verify that the host has access to the required data.
5.1.3 API tables supported by the API referral application
The API tables listed here are supported by the API referral application.
Table 5-1: API tables, process fluids, measurement units, and default reference values
Referred
Process fluid
density
Generalized crude
5A
and JP4
23A
53A
Generalized
5B
products (gasoline,
jet fuel, aviation fuel,
56
API tables
(1)
(calculations)
Referred density
CTL or
(API): unit and
(2)
(3) (4)
CTPL
range
6A
Unit: °API
Range: 0 to 100 °API
24A
Unit: SGU
Range: 0.6110 to
1.0760 SGU
54A
Unit: kg/m
Range: 610 to
1075 kg/m³
6B
Unit: °API
Range: 0 to 85 °API
Setup
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Click Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
Default
reference
temp
60 °F
60 °F
3
15 °C
60 °F
Micro Motion Model 5700 with Ethernet Transmitters
Configuration and Use Manual
MMI-20029769
Default
reference
pressure
API standard
0 psi (g)
API MPMS 11.1
0 psi (g)
0 kPa (g)
0 psi (g)
API MPMS 11.1