Integrate the meter with the control system
March 2019
Effect of mA Output Direction on an mA Output
mA Output Direction affects how the transmitter reports flow values via an mA Output. An mA Output is
affected by mA Output Direction only if mA Output Source is set to a flow variable.
The effect of mA Output Direction depends on the setting of Lower Range Value (LRV).
If Lower Range Value = 0, see
If Lower Range Value > 0, see
If Lower Range Value < 0, see
Figure 8-1: Effect of mA Output Direction on an mA Output: Lower Range Value = 0
mA Output Direction = Normal
20
12
4
-x
Reverse flow
Lower Range Value = 0
Upper Range Value = x
114
Figure
8-1.
Figure 8-1
and adapt the chart.
Figure
8-2.
mA Output Direction = Absolute Value
0
x
Forward flow
Configuration and Use Manual
20
12
4
-x
0
Reverse flow
Micro Motion Model 5700 with Ethernet Transmitters
MMI-20029769
x
Forward flow