Option
Fix at Fault
5. Click Start and wait for the test to complete.
Tip
At any time during the process, you can abort the test. If the outputs were fixed, they will return to
normal behavior.
When SMV is licensed, results for the test are stored in the transmitter memory, along with results for tests
performed with any other tool. These results are also stored in the ProLink III database. You can view and
use these results in any tool-based trending and reporting functions.
When SMV is not licensed, results from previous tests are saved in a database, but not accessible. A Pass/
Fail indication is all that is provided for the current test.
Postrequisites
View the results and take any appropriate actions.
Run an SMV test using a web browser
Procedure
1. Read the Prerequisites in
2. Choose Service Tools > Smart Meter Verification > Run Meter Verification.
3. Select the desired output behavior.
Option
Continue Measuring During the test, all outputs will continue to report their assigned process
Fix at Last
Measured Value
Fix at Fault
4. Click Start and wait for the test to complete.
Tip
At any time during the process, you can abort the test. If the outputs were fixed, they will return to
normal behavior.
When SMV is licensed, results for the test are stored in the transmitter memory, along with results for tests
performed with any other tool. These results are also stored in the ProLink III database. You can view and
use these results in any tool-based trending and reporting functions.
When SMV is not licensed, results from previous tests are saved in a database, but not accessible. A Pass/
Fail indication is all that is provided for the current test.
Configuration and Use Manual
Description
During the test, all outputs will go to their configured fault action. The test will
run for approximately 140 seconds.
Use Smart Meter Verification (SMV)
Description
variables. The test will run for approximately 90 seconds.
During the test, all outputs will report the last measured value of their assigned
process variable. The test will run for approximately 140 seconds.
During the test, all outputs will go to their configured fault action. The test will
run for approximately 140 seconds.