14.1.3 Set up SMV automatic execution
You can set up and run a single test at a user-defined future time. You can also set up and run tests
automatically on a regular schedule.
Automatic execution of SMV is managed from the transmitter. You do not need a connection from an
external configuration tool.
Tip
The time between test runs must be between 1 and 1000 hours. The time to the first test run can be any
positive floating number.
Important
Only the 20 most recent SMV results are stored. For SMV Basic versions, the last automatic execution result
can be viewed, while all 20 results can be viewed with SMV licensed. To view or chart these results using an
external tool, you must upload them from the transmitter.
Set up SMV automatic execution using the display
Procedure
1. Choose Menu > Service Tools > Verification & Calibration > Smart Meter Verification > Schedule
SMV.
2. To schedule a single test:
a) Set Hours to 1st Run to the number of hours to elapse before the test is run.
b) Set Hours Between to 0 .
3. To schedule a recurring execution:
a) Set Specify Time Until Next Run to the number of days, hours, and minutes to elapse before the
first test is run.
b) Set Specify Time Between Recurring Runs to the number of days, hours, and minutes to elapse
between runs.
4. To disable scheduled execution:
a) Set Specify Time Until Next Run to 0 days, 0 hours, and 0 minutes.
b) Set Specify Time Between Recurring Runs to 0 days, 0 hours, and 0 minutes.
Set up SMV automatic execution using ProLink III
Procedure
1. Select one of the following paths to access the SMV scheduler .
Choose Device Tools > Diagnostics > Meter Verification > Meter Verification Scheduler.
Choose Smart Meter Verification Overview > Tools > Schedule Smart Meter Verification.
2. To schedule a single test:
a) Set Specify Time Until Next Run to the number of days, hours, and minutes to elapse before the
test is run.