Emerson Micro Motion 5700 Configuration And Use Manual page 236

With ethernet transmitters
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
March 2019
17.21.6 Perform a Frequency Output loop test using ProLink III
Prerequisites
Before performing a Frequency Output loop test, configure Channel C as a Frequency Output.
Follow appropriate procedures to ensure that loop testing will not interfere with existing measurement and
control loops.
Procedure
1. Choose Device Tools > Diagnostics > Testingand select the Frequency Output to test.
2. Enter the Frequency Output value in Fix to.
3. Select Fix FO.
4. Read the frequency signal at the receiving device and compare it to the transmitter output.
5. Select UnFix FO.
17.21.7 Perform a Discrete Output loop tests using ProLink III
Prerequisites
Before performing a Discrete Output loop test, configure Channel C as a Discrete Output.
Follow appropriate procedures to ensure that loop testing will not interfere with existing measurement and
control loops.
Procedure
1. Choose Device Tools > Diagnostics > Testing > Discrete Output Test.
2. If your transmitter is configured for two or more Discrete Outputs, determine which Discrete Output
you want to test.
3. Set Fix To: to ON.
4. Verify the signal at the receiving device.
5. Set Fix To: to OFF.
6. Verify the signal at the receiving device.
7. Select UnFix.
Postrequisites
If the Discrete Output readings are reversed, check the setting of Discrete Output Polarity.
17.21.8 Perform a Discrete Input loop test using ProLink III
Prerequisites
Before performing a Discrete Input loop test, configure Channel C for a Discrete Input.
Follow appropriate procedures to ensure that loop testing will not interfere with existing measurement and
control loops.
Procedure
1. Set the remote input device to ON.
236