Recommended actions
Disable sensor simulation.
17.12.37 [138] TBR Active
Alert
Process Aberration
Cause
Drive gain is above the configured threshold and transient bubble remediation is active.
Recommended actions
Check for air entrainment, tube fouling, flashing, or tube damage.
17.12.38 Bad IP network settings
Alert
Configuration Error
Cause
IF DHCP is enabled, the DHCP server could not be found.
If DHCP is disabled, there is a mismatch between the IP address, Subnet mask and default gateway
settings.
Recommended actions
1. Correct the network settings on the transmitter and/or the PC.
2. If the transmitter is installed on an Ethernet network, contact your network administrator to determine
the correct network settings.
17.12.39 Batcher Not Configured
Alert
Configuration Error
Cause
One or more of the following has occurred:
The batcher application is disabled
No flow source has been configured for the batcher application
The batch target is 0
No discrete output is assigned to batch control.
Recommended actions
Verify the configuration of the batcher application and related parameters.
Configuration and Use Manual
Troubleshooting