Measurement support
March 2019
4. Solve for A (slope).
5. Solve for B (offset), using the calculated slope and one set of values.
6. Enter the results as the trim slope and the trim offset.
Using ProLink III: Choose Device Tools > Configuration > Process Measurement > Concentration
Measurement, set Matrix Being Configured to your matrix, and enter Trim Slope and Trim Offset.
Using a Web browser: Choose Configure
7. Take another concentration reading from the meter, and compare it to the laboratory value.
If the two values are acceptably close, the trim is complete.
If the two values are not acceptably close, repeat this procedure.
Calculating the trim slope and the trim offset
Comparison 1
Comparison 2
Populate the equations:
Solve for A:
Solve for B:
Concentration slope (A): 1.00029
Concentration offset (B): 0.00551
176
Laboratory value
Meter value
Laboratory value
Meter value
50 = A × 49.98 + B
16 = A × 15.99 + B
50.00 − 16.00 = 34.00
49.98 − 15.99 = 39.99
34 = A × 33.99
A = 1.00029
50.00 = 1.00029 × 49.98 + B
50.00 = 49.99449 + B
B = 0.00551
Configuration and Use Manual
50.00%
49.98%
16.00%
15.99%
Micro Motion Model 5700 with Ethernet Transmitters
MMI-20029769