Using the SatSeconds™ Alarm Management System
4.5.4
The SatSeconds™ feature monitors both degree and duration of desaturation as an index of desat-
uration severity. Use the appropriate SatSeconds™ setting—Off, 10, 25, 50, or 100—to distinguish
clinically significant events from minor and brief desaturations that may result in nuisance alarms.
The SatSeconds™ alarm management system calculates the duration of the event multiplied by the
number of percentage points that saturation falls outside of the saturation alarm threshold. With
SatSeconds™, an alarm is triggered only when the SatSeconds™ value is reached.
For more information about SatSeconds™, reference
page 10-6.
To set SatSeconds™
Press the Menu button.
1.
Press the Down button to highlight the Alarm Limits menu item and press OK to select it.
2.
On the Alarm Limits menu, press the Down button to highlight SatSeconds.
3.
4-12
Figure 4-7.  High SpO
for adult and neonate limit options.
4-7
shows the High SpO
Setting
2
Menu Structure and Factory Defaults
SatSeconds™ Alarm Management
setting
2