Biotek Synergy HT Operator's Manual page 83

Multi-mode microplate reader
Hide thumbs Also See for Synergy HT:
Table of Contents

Advertisement

Reader:
Basecode:
Date and Time:
User:
Company:
Comments:
Test Results
Operator ID:______________________________________________________________
Notes:____________________________________________________________________
7090202
Version 2.23
Bias current offset
Offset voltage
750V measurement
750V noise
500V measurement
500V noise
Lambda: 200
Gain:
Channel:
Ref
Air:
14235
Dark:
9869
Delta:
4366
Lambda: 340
Gain:
Channel:
Ref
Air:
12954
Dark:
9871
Delta:
3083
Lambda: 405
Gain:
Channel:
Ref
Air:
12755
Dark:
9871
Delta:
2884
Lambda: 550
Gain:
Channel:
Ref
Air:
12783
Dark:
9869
Delta:
2914
Lambda: 630
Gain:
Channel:
Ref
Air:
12705
Dark:
9868
Delta:
2837
Lambda: 999
Gain:
Channel:
Ref
Air:
13084
Dark:
9865
Delta:
3219
Synergy HT Operator's Manual
Gen5 System Test Report
Synergy
(Serial Number: 128787
P/N 7090202
07/08/2012 10:12:58 AM
Administrator
BioTek
System Test run during the IQ
SYSTEM SELF TEST
128787
-0.9 counts
1541 counts
90.1 counts
66 counts
4.6 counts
3 counts
1.80
Resets: 1
1
39595
9889
29706
1.95
Resets: 2
1
39951
9885
30066
1.61
Resets: 2
1
38902
9883
29019
1.63
Resets: 1
1
39796
9887
29909
2.81
Resets: 1
1
39020
9896
29124
4.49
Resets: 1
1
39360
9911
29449
Figure 29: Sample output for the System Test
)
(v2.24)
1111 1110
PASS
PASS
PASS
59
System Test |

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Synergy HT and is the answer not in the manual?

Questions and answers

Luqman Babeer
April 25, 2025

I Connected ELx800 with Gen5 3.12 .When i check the connection it is succeeded because i was make plate in and plate out successfully and message appear that connected successfully .But when i need to run plate or tests the message appear (Reader Failed to acknowledge request) or when i need to make validation it will be fail . I don't know how can I resolve this issue. I sent the errors photo here Thank you

User image 680b7dad96763
1 comments:
Mr. Anderson
May 12, 2025

To troubleshoot the "Reader Failed to acknowledge request" error on the Biotek Synergy HT when connected to ELx800 with Gen5 3.12, follow these steps:

1. Ensure the reader is connected to the power supply and turned on.
2. Verify the communication cable is firmly attached to both the reader and the computer.
3. Confirm the correct Reader Type is selected in Gen5.
4. Try using a different COM port.
5. If using a USB cable, ensure the driver software is installed.

If the issue persists, contact BioTek’s Technical Assistance Center.

This answer is automatically generated

Subscribe to Our Youtube Channel

Table of Contents

Save PDF