Appendix A: Decontamination
|
8.
Reassemble the instrument as necessary.
9.
If the dispense module is installed, detach the outlet tubes from the rear panel
of the instrument. If it is not installed, attach just the dispense module's
communication cable to the instrument. Remove the supply bottles and their
holders.
10.
Perform the procedures described below through page 130 to decontaminate
the fluid lines in the dispense module, the internal tubing and injector heads,
and the tip priming trough and priming plate.

Decontaminate the Fluid Lines

1.
Place a beaker with 20 mL of 0.5% sodium hypochlorite solution or 70%
isopropyl alcohol near
Place the SYRINGE 1
2.
3.
If you have not already done so, detach the dispense module's
from the instrument's rear panel. Place the ends of the outlet tubes in an empty
beaker and set the beaker on the work surface.
4.
Launch Gen5, select
Dispenser
5.
Select
Dispenser 1
dispense
Place the
6.
presence).
7.
Run
two
8.
Pause for
9.
Remove the
10.
From the
11.
Run
one
12. Empty
Important! If sodium hypochlorite (bleach) was used, perform
13.
Lines
Otherwise, (or after performing the Rinse procedure), repeat steps 1–13 for
SYRINGE 2
SYRINGE 1
inlet tube
System > Reader/Instrument Control,
tab.
, enter a
Rate.
on the carrier (it is not used, but the reader requires its
priming plate
prime cycles, for a total of 10000 µL.
20 to 30 minutes
inlet tube
from the beaker of disinfectant solution.
Reader Control
dialog, change the
prime cycle, to flush the disinfectant out of the fluid lines.
the beaker containing the outlet tubes. Put the tubes back in.
on the next page.
/
Dispenser 2
on the dispense module.
in the beaker.
Volume
of
5000 µL
, and keep the default
to allow the solution to disinfect the tubing.
Volume
.
outlet tubes
and click the
to
1000 µL.
Rinse the Fluid
BioTek Instruments, Inc.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Synergy HT and is the answer not in the manual?

Questions and answers

Luqman Babeer
April 25, 2025

I Connected ELx800 with Gen5 3.12 .When i check the connection it is succeeded because i was make plate in and plate out successfully and message appear that connected successfully .But when i need to run plate or tests the message appear (Reader Failed to acknowledge request) or when i need to make validation it will be fail . I don't know how can I resolve this issue. I sent the errors photo here Thank you

User image 680b7dad96763
1 comments:
Mr. Anderson
May 12, 2025

To troubleshoot the "Reader Failed to acknowledge request" error on the Biotek Synergy HT when connected to ELx800 with Gen5 3.12, follow these steps:

1. Ensure the reader is connected to the power supply and turned on.
2. Verify the communication cable is firmly attached to both the reader and the computer.
3. Confirm the correct Reader Type is selected in Gen5.
4. Try using a different COM port.
5. If using a USB cable, ensure the driver software is installed.

If the issue persists, contact BioTek’s Technical Assistance Center.