Code
Measurement channel correction outside limits
0F11 – 0F16
See criteria in text below. See Delta (Air – Dark) on self-test.
This error can indicate one of the following error scenarios:
• During a system test, the Delta was < 8000 for one of the wavelengths.
• During a spectral scan or blank read, the flash-on value minus flash-off was
< 8000 normal mode, or < 8000 * calibrated resets / 8 for sweep mode.
Probable Causes:
• Lamp, alignment, or lamp power supply failure.
• Absorbance channel analog PCB.
• Order-sorting filter wheel is jammed, not aligning the correct bandpass filter with the light
path and white light is passing through, or the filter is degraded and is not passing enough
light energy or is blocking the light.
• Damaged optic spray.
1000
Necessary configuration data missing
(This error code indicates old basecode. Recommend upgrading.)
Probable Causes:
• The flash memory on the 7080400 PCB is defective or corrupt. The basecode software
and/or assays may need to be re-downloaded.
Failed configuration checksum test
1100
This error indicates that during self-test or at the end of a plate read, the checksums calculated
for configuration flash memory page 0 do not match the saved checksums.
Probable Causes:
• The flash memory on the 7080400 PCB is defective or corrupt. The basecode software may
need to be re-downloaded.
Failed configuration checksum test
1101
This error indicates that during self-test or at the end of a plate read, the checksums calculated
for configuration flash memory page 1 do not match the saved checksums.
Probable Causes:
• The flash memory on the 7080400 PCB is defective or corrupt. The basecode software may
need to be re-downloaded.
Autocalibration data missing for the Fluorescence top/bottom and absorbance reads
1200
This error indicates that no autocal data for any of the three read locations (bottom probe, top
probe, absorbance).
Probable Causes:
• The 7080400 PCB was changed and the Flash memory does not have the calibration values
loaded. Performing the autocalibration procedure will correct this.
Synergy HT Operator's Manual
Description and Probable Causes:
157
Error Codes |
Need help?
Do you have a question about the Synergy HT and is the answer not in the manual?