Error Codes; Fatal Errors - Biotek Synergy HT Operator's Manual

Multi-mode microplate reader
Hide thumbs Also See for Synergy HT:
Table of Contents

Advertisement

Fatal Errors

Fatal errors indicate conditions that require immediate attention. If a fatal error is
displayed, contact BioTek's Technical Assistance Center for further instructions.
Code
Task
A100
control
A200
Read already in progress
Motor not available
A300
This error indicates that a motor is not available but it does not identify which motor was
requested.
A301
Real-time clock not available
A302
Display device not available
A304
T0 device not available
Timer 0, in addition to being used by the motors, is also used in microsecond timing for time-
resolved measurements.
A400
Failed code checksum test on power-up
DR steps alloc/free error <assay number>
A500
(This error indicates old basecode; recommend upgrading)
A502
24V power dropped below safe level
A600
Data flash write timed out
A700
Data flash read back did not match write
A800
Code flash write timed out
A900
Memory allocation heap corrupted
Absorbance A/D converter did not see the Ready signal
AA01
(This error indicates old basecode; recommend upgrading)
Fluorescence A/D Converter did not see the Ready signal
AA02
(This error indicates old basecode; recommend upgrading)
Probable Causes:
• PMT detector PCB (AA02)
• Analog PCB (AA01)
• Motor/Power PCB (AA01 or AA02)
Synergy HT Operator's Manual
Description
block not available
137