Absorbance Plate Test; Test Plate Certificates; Define Absorbance Test Plate Parameters - Biotek Synergy HT Operator's Manual

Multi-mode microplate reader
Hide thumbs Also See for Synergy HT:
Table of Contents

Advertisement

62
Chapter 4: Instrument Qualification
|

Absorbance Plate Test

This test uses BioTek's
alignment; optical density accuracy, linearity, and repeatability; and wavelength accuracy
of the Synergy HT. The Absorbance Plate Test compares the reader's optical density and
wavelength measurements to NIST-traceable values.
An alternate method that may be used to determine accuracy,
linearity, and repeatability is Liquid Test 2, described on page 71.

Test Plate Certificates

To run this test, you will need BioTek's Absorbance Test Plate (PN 7260522), with its
accompanying data sheet.
The Absorbance OD Standards section contains NIST-traceable standard OD
values for the filters at several different wavelengths. We recommend testing at six
wavelengths—those at or close to the wavelengths used in your assays.
The Wavelength Accuracy Standards section contains Expected Peak wavelength
values for the filter in position C6 on the plate. Each value has a valid test range
associated with it. For example, an Expected Peak value may be 586 nm with
tolerance values of –6/+4 (or a test range of 580 to 590 nm).
The instructions provided below and on the following page are
guidelines. Refer to the Gen5 Help system for more information.

Define Absorbance Test Plate Parameters

Gen5 version 1.07 and earlier users only: The Gen5 Reader
Diagnostics Utility must be installed to run the Absorbance Plate Test.
Before the Absorbance Plate Test can be performed, the standard OD
values and the peak wavelength value(s) must be entered into Gen5.
You'll enter and save these values once initially, and then update them
each time the test plate is recertified by BioTek (typically annually).
1.
Obtain the data sheet that came with the Test Plate.
2.
Start Gen5 and select
Plates
3.
Click
Add
4.
Select the appropriate Plate Type and enter the plate's Serial Number.
5.
Enter the Last Certification and Next Certification dates from the calibration
sticker on the Test Plate.
Absorbance Test Plate
System > Diagnostics > Test Plates > Add/Modify
.
.
(PN 7260522) to confirm the mechanical
BioTek Instruments, Inc.