Reader:
Basecode:
Date and Time:
Absorbance Plate:
Last Plate Certification:
Next Plate Certification Due: January 2013
User:
Comments:
Peak Absorbance Results
Well
C6
Reference
586
Tolerance
3
Read
587
Result
PASS
Alignment Results
Wells
A1
Read
0.001
Tolerance
0.015
Result
PASS
Wavelength = 405 nm
Accuracy Results
Wells
C1
Reference
0.147
Min Limit
0.124
Max Limit
0.170
Read 1
0.144
Result
PASS
Repeatability Results
Wells
C1
Read 1
0.144
Min Limit
0.138
Max Limit
0.150
Read 2
0.144
Result
PASS
Wavelength = 630 nm
Accuracy Results
Wells
C1
Reference
0.136
Min Limit
0.113
Max Limit
0.159
Read 1
0.134
Result
PASS
Synergy HT Operator's Manual
Absorbance Test Plate Results
Synergy
P/N 7090202
07/08/2012 03:03:36 PM
7 Filter Test Plate (P/N 7260522) - S/N 161259
January 2012
Administrator
Test performed during Initial OQ
A12
0.002
0.015
PASS
E2
0.618
0.586
0.650
0.615
PASS
E2
0.615
0.604
0.626
0.615
PASS
E2
0.568
0.537
0.599
0.566
PASS
Figure 30: Sample output for the Absorbance Plate Test
(Serial Number: 128787)
(v2.24)
H1
H12
0.001
0.002
0.015
0.015
PASS
PASS
G3
H6
1.133
1.701
1.090
1.647
1.176
1.755
1.128
1.696
PASS
PASS
G3
H6
1.128
1.696
1.112
1.674
1.144
1.718
1.128
1.695
PASS
PASS
G3
H6
1.040
1.560
0.999
1.509
1.081
1.611
1.037
1.557
PASS
PASS
Absorbance Plate Test |
F5
D4
2.279
2.945
2.168
2.807
2.390
3.083
2.284
2.908
PASS
PASS
F5
D4
2.284
2.908
2.210
2.816
2.358
3.000
2.285
2.903
PASS
PASS
F5
D4
1.865
2.400
1.808
2.284
1.922
2.516
1.866
2.385
PASS
PASS
65
Need help?