Cleaning Exposed Surfaces
Exposed surfaces may be cleaned (not decontaminated) with a cloth moistened
(not soaked) with water or water and a mild detergent. You will need:
•
Deionized or distilled water
•
Clean, lint-free cotton cloths
•
Mild detergent (optional)
To clean the exposed surfaces:
1.
Turn off and unplug the instrument.
2.
Moisten a clean cotton cloth with water, or with water and mild detergent.
Do not soak the cloth
3.
Wipe the plate carrier and all exposed surfaces of the instrument.
4.
Wipe all exposed surfaces of the dispense module (if used).
5.
If detergent was used, wipe all surfaces with a cloth moistened with water.
6.
Use a clean, dry cloth to dry all wet surfaces.
7.
Reassemble the instrument as necessary.
Models with injectors: If the Tip Priming Trough overflows, wipe the
carrier and the surface beneath the carrier with a dry cotton cloth. If
overflow is significant, you may have to remove the shroud of the
instrument to better access the surface beneath the carrier.
See page 106 for instructions on removing the shroud.
See page 120 for instructions for cleaning the surface beneath the
carrier.
Synergy HT Operator's Manual
Important!
Turn off and unplug the instrument for all
cleaning operations.
Important!
Do not immerse the instrument, spray it with
liquid, or use a "wet" cloth. Do not allow the cleaning
solution to run into the interior of the instrument. If this
happens, contact BioTek's Service Department.
.
Cleaning Exposed Surfaces |
99
Need help?
Do you have a question about the Synergy HT and is the answer not in the manual?
Questions and answers
I Connected ELx800 with Gen5 3.12 .When i check the connection it is succeeded because i was make plate in and plate out successfully and message appear that connected successfully .But when i need to run plate or tests the message appear (Reader Failed to acknowledge request) or when i need to make validation it will be fail . I don't know how can I resolve this issue. I sent the errors photo here Thank you
To troubleshoot the "Reader Failed to acknowledge request" error on the Biotek Synergy HT when connected to ELx800 with Gen5 3.12, follow these steps:
1. Ensure the reader is connected to the power supply and turned on.
2. Verify the communication cable is firmly attached to both the reader and the computer.
3. Confirm the correct Reader Type is selected in Gen5.
4. Try using a different COM port.
5. If using a USB cable, ensure the driver software is installed.
If the issue persists, contact BioTek’s Technical Assistance Center.
This answer is automatically generated