Pipette Map
Seal the plates with foil or store them in black polyethylene bags until use.
If the base of a plate is touched, clean the entire base with alcohol (95%
ethanol) and then wipe with a lint-free cloth. Before placing a plate in the
instrument, blow the bottom of the plate with an aerosol duster.
For the
Corners
•
Pipette 200 µL of the
H1-H3, and H10-H12.
•
(Optional for plates with black sidewalls) Pipette 200 µL of buffer in the
wells surrounding the 3.3 nM wells ('CBUF' in the grid below).
For the
Sensitivity/Linearity
•
Use a multi-channel pipette with just four tips installed.
•
Pipette 150 µL of buffer into wells C2-F5 (NOT column 1) and C10-F12.
•
Pipette 150 µL of the
•
Pipette 150 µL of the
Mix the wells using the pipette. Do not discard the tips.
•
Aspirate 150 µL from wells C2-F2 and dispense into wells C3-F3.
Mix the wells using the pipette. Do not discard the tips.
•
Aspirate 150 µL from wells C3-F3 and dispense into wells C4-F4.
Mix the wells using the pipette. Do not discard the tips.
•
Aspirate 150 µL from wells C4-F4 and dispense into wells C5-F5.
Mix the wells using the pipette.
•
Aspirate 150 µL from wells C5-F5 and discard.
1
2
3.3
3.3
A
nM
nM
B
CBUF
CBUF
1.0
0.5
C
nM
nM
1.0
0.5
D
nM
nM
1.0
0.5
E
nM
nM
1.0
0.5
F
nM
nM
G
CBUF
CBUF
3.3
3.3
H
nM
nM
Synergy HT Operator's Manual
test:
3.3 nM SF
tests:
1 nM SF
1 nM SF
3
4
5
3.3
CBUF
nM
CBUF
CBUF
0.25
0.125
0.062
nM
nM
nM
0.25
0.125
0.062
nM
nM
nM
0.25
0.125
0.062
nM
nM
nM
0.25
0.125
0.062
nM
nM
nM
CBUF
CBUF
3.3
CBUF
nM
Fluorescence Tests |
solution into wells A1-A3, A10-A12,
solution into wells C1-F1. Discard the tips.
solution into wells C2-F2.
6
7
8
CBUF
CBUF
CBUF
CBUF
9
10
11
12
3.3
3.3
3.3
nM
nM
nM
CBUF
CBUF
CBUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
CBUF
CBUF
CBUF
3.3
3.3
3.3
nM
nM
nM
81
Need help?
Do you have a question about the Synergy HT and is the answer not in the manual?
Questions and answers
I Connected ELx800 with Gen5 3.12 .When i check the connection it is succeeded because i was make plate in and plate out successfully and message appear that connected successfully .But when i need to run plate or tests the message appear (Reader Failed to acknowledge request) or when i need to make validation it will be fail . I don't know how can I resolve this issue. I sent the errors photo here Thank you
To troubleshoot the "Reader Failed to acknowledge request" error on the Biotek Synergy HT when connected to ELx800 with Gen5 3.12, follow these steps:
1. Ensure the reader is connected to the power supply and turned on.
2. Verify the communication cable is firmly attached to both the reader and the computer.
3. Confirm the correct Reader Type is selected in Gen5.
4. Try using a different COM port.
5. If using a USB cable, ensure the driver software is installed.
If the issue persists, contact BioTek’s Technical Assistance Center.