11 Maintenance
11.1 General instructions
11.2 Intervals
Part concerned
Device
Battery
Disposable hose system
Reusable hose system
FlowCheck sensor
Device input filter
Hygiene filter
Maintenance, safety checks, inspections and repairs must only be
carried out by the manufacturer or a technician specifically
authorized by the manufacturer.
Interval
Maintenance and safety check
every 2 years
Maintenance-free
When stored in the device: Charge every 3 months.
When not stored in the device: Charge every 5 months.
Recommendation: Replace battery after 2 years.
Maintenance-free
Maintenance every 2 years
Following a prompt during the
function check
Recommendation: Replace
FlowCheck sensor after 2 years.
Following a prompt during the
function check (every 6 months or
after 24 ventilation hours in Air
Mix mode)
Following a prompt during the
function check (every 6 months or
after 24 ventilation hours in Air
Mix mode)
or
after each transportation of an
infected and ventilated patient
Maintenance by
Manufacturer or a technician
specifically authorized by the
manufacturer
User/operator
"11.4 Maintaining the reusable
hose system", page 207)
User/operator
User/operator
"11.6 Replacing the device input
filter", page 209)
User/operator
"11.5 Replacing the hygiene
filter", page 208)
MEDUMAT Standard
11 Maintenance
(see
(see
(see
2
EN
205