2 Safety
Notice
2.1.2 Power supply
Warning
10
EN
MEDUMAT Standard
Risk of injury as a result of pressure variations during use in
combination with devices from the WEINMANN Emergency
MODUL range!
If the device is used together with devices from the WEINMANN
Emergency MODUL range, the flow used by devices from the
WEINMANN Emergency MODUL series may cause pressure
variations in the device. This can injure the patient.
Only use the device and devices from the WEINMANN
Emergency MODUL range in combinations approved by
WEINMANN Emergency.
Delay in treatment due to interference caused by electric and
magnetic fields!
Electric and magnetic fields may interfere with device functioning,
and delay treatment.
Maintain separation distances between the device and mobile
telephones, radio units and X-ray apparatus.
Damage to the device caused by ingress of liquids!
The device is rated IP54 (splash-proof). This only applies when the
battery is located in the battery compartment, the SD card
compartment is closed, there is a filter in the filter compartment,
and there is a patient hose system connected. Ingress of liquids
may damage the device, components and accessories.
Do not immerse the device, components, or accessories in
liquids.
Clean the battery compartment carefully so that no liquids
enter the device.
Risk of injury due to missing, flat, or defective battery!
A missing, flat or defective battery prevents treatment.
Only operate the device with a charged battery.
Keep an alternative ventilation unit at the ready.
Treatment prevented by defective power cord or power
supply!
A defective power cord or power supply prevents the battery in
the device from charging and thus impairs the operational
readiness of the device.
Inspect the power cord and power supply regularly.
Only operate the device with a charged battery.
Keep an alternative ventilation unit at the ready.
2