Simv + Asb Mode - Weinmann MEDUMAT Standard2 Instructions For Use Manual

Ventilator
Hide thumbs Also See for MEDUMAT Standard2:
Table of Contents

Advertisement

The SIMV mode is used for volume-controlled ventilation with a
fixed mandatory minute volume. The patient can breathe
spontaneously between the mandatory mechanical breaths and
thereby increase the minute volume. During spontaneous
respiration, the mandatory mechanical breath is synchronized with
the patient's breathing. The mandatory minute volume and the
mandatory respiration rate remain unchanged. The set maximum
pressure limitation (pMax) ensures the safety of the patient.

7.3.9 SIMV + ASB mode

Description
Abbreviation
Long form
Type
Requirement
Ventilation parameters
Left-hand navigation knob
Central navigation knob
Right-hand navigation knob
7 Description of the modes
SIMV + ASB
Synchronized Intermittent Mandatory
Ventilation + Assisted Spontaneous
Breathing
Volume-controlled
SIMV option is activated
Flow measurement + ASB option
is activated
Vt
Freq.
PEEP
pMax
Δ pASB
InTr
ExTr (does not appear if at least
one pressure-controlled ventilation
mode is activated)
I:E
Emergency mode
MEDUMAT Standard
2
EN
151