Example
(1)
Source: TRAUB, S.L.; JOHNSON, C.E.: Comparison of methods of estimating creatinine clearance in children. In:
American journal of hospital pharmacy 37, 1980, No.2, pp. 195–201.
(2)
Source: DEVINE, Ben J. Gentamicin therapy. The Annals of Pharmacotherapy, 1974, Vol. 8, No. 11, pp. 650-655

14.3 Exported log files

File name
debug
status
fcheck
mission logs

14.3.1 Recorded function checks

Patient, male, height 185 cm
Setting for Vt/kg KG = 6 ml/kg
 IBW = 50 + 2.3 x [185 cm/2.54 - 60] = 79.51 kg
 Vt = 80 kg x 6 ml/kg = 450 ml
If you have exported log files to an SD card
Export",
page 121), you will find the following files on the SD card:
Description
Supports communication in the event of servicing.
Supports communication in the event of servicing.
Record of the function checks which have been performed
Detailed recording of session data
In the file fcheck, the function checks which have been performed
are saved along with the date, time and their results. This
information helps you with documentation within the scope of
your quality management system. You can open the file fcheck
with a spreadsheet program (e.g., Microsoft
In the column result, you will find the overall result of a function
check (ok = passed, failed = not passed). A function check is failed
if a test is not passed.
(see "6.3.4 Import /
®
Excel
MEDUMAT Standard
14 Appendix
80 kg
(see 14.3.1,
p. 235).
®
).
2
EN
235

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 29500

Table of Contents