6 Operator menu

6.3.8 Presets patient

128
EN
MEDUMAT Standard
6-9 Presets patient submenu
In the Presets patient submenu, you can determine which presets
are assigned to the ventilation parameters of the different patient
groups:
Risk of injury from different alarm presets in the same or
similar devices!
Different alarm presets in the same or similar devices in different
application areas can confuse the user and result in injury to the
patient.
 Select the same alarm presets in the same or similar devices.
Risk of injury from deactivated alarms in the additional
functions CPR and RSI!
Through deactivation in the operator menu, no alarms for flow
and CO
monitoring are output in the additional functions CPR
2
and RSI. This can injure the patient.
 Notify the user if the alarms in the additional functions CPR and
RSI have been deactivated.
 Observe the alarm limits during ventilation in the additional
functions CPR and RSI.
2